CMS8000 CONTEC ICU CCU Vital Signs Patient…: Uncontrolled Resource Consumption
CVE-2022-38100 is a high-severity uncontrolled resource consumption vulnerability in CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor from Contec Health, with a CVSS 3.1 base score of 7.5 out of 10. It can be exploited remotely over the network by an unauthenticated attacker, without any action from a user. Affected CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor versions: All. The record does not name a fixed version, so check the vendor advisory. There is no confirmed exploitation so far. EPSS estimates a 1.05% chance of exploitation within 30 days, higher than 65% of scored CVEs.
High severity with no evidence of exploitation so far.
What to do
- Check which version of CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor you run against the affected versions below, and install the vendor's latest release.
- Until you can update, restrict network access to the affected service so it is not reachable from untrusted networks.
Affected products and versions
| Vendor | Product | Affected versions | Fixed in |
|---|---|---|---|
| Contec Health | CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor | All | Not stated |
| Contechealth | Cms8000 Firmware | Not specified in the record | Not stated |
Version data comes from the organisation that assigned the CVE. Confirm against the vendor advisory before deciding you are unaffected.
How the severity score breaks down
- Attack vector
- NetworkCan be exploited remotely over a network, including the internet.
- Attack complexity
- LowNo special conditions are needed; the attack works reliably.
- Privileges required
- NoneNo account or login is needed.
- User interaction
- NoneNo action by a victim is required.
- Scope
- UnchangedImpact stays within the vulnerable component.
- Confidentiality
- NoneNo data is disclosed.
- Integrity
- NoneNothing can be modified.
- Availability
- HighThe component can be taken down completely.
Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Scored by NVD.
| Scoring system | Score | Severity | Scored by |
|---|---|---|---|
| CVSS 3.1 | 7.5 | High | CNA (hq.dhs.gov) |
| CVSS 3.1 | 7.5 | High | NVD |
Official description
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
Source: the CVE record as published by its numbering authority, via NVD and CVE.org.
The underlying weakness: Uncontrolled Resource Consumption
The product does not properly control the allocation and maintenance of a limited resource.
How developers prevent this class of flaw
- Ensure that protocols have specific limits of scale placed on them.
- Ensure that all failures in resource allocation place the system into a safe posture.
More Uncontrolled Resource Consumption vulnerabilities (CWE-400)
Timeline
- CVE record published.
- CISA published its SSVC assessment (exploitation: none, automatable: yes, technical impact: partial).
- Record last changed materially (score, affected products, status or references).
References
Patches and release notes
- cisa.gov/uscert/ics/advisories/icsma-22-244-01Mitigation, Third Party Advisory, US Government Resource
Recent Uncontrolled Resource Consumption vulnerabilities
- 7.5 High severity
- 7.5 High severity
- 7.5 High severity
- 7.5 High severity
- 7.5 High severity
- 7.5 High severity
Spotted an error? Tell us.