CyberSecBulletin
CVE-2023-1968NVD status: Modified

iScan Control Software: Binding to an Unrestricted IP Address

CVE-2023-1968 is a high-severity binding to an unrestricted IP address vulnerability in iScan Control Software from Illumina, with a CVSS 3.1 base score of 7.5 out of 10. It can be exploited remotely over the network by an unauthenticated attacker, without any action from a user. Affected iScan Control Software versions: 4.0.0; 4.0.5. The record does not name a fixed version, so check the vendor advisory. There is no confirmed exploitation so far. EPSS estimates a 1.81% chance of exploitation within 30 days, higher than 79% of scored CVEs.

Next patch cycle

High severity with no evidence of exploitation so far.

What to do

  1. Check which version of iScan Control Software you run against the affected versions below, and install the vendor's latest release.
  2. Read the advisory on support.illumina.com for workarounds and exact fixed builds.
  3. Until you can update, restrict network access to the affected service so it is not reachable from untrusted networks.

Affected products and versions

VendorProductAffected versionsFixed in
IlluminaiScan Control Software4.0.0; 4.0.5Not stated
IlluminaMiniSeq Control Software2.0Not stated
IlluminaiSeq 100All versionsNot stated
IlluminaMiSeq Control Software4.0 (RUO Mode)Not stated
IlluminaMiSeqDx Operating Software4.0.1Not stated
IlluminaNextSeq 500/550 Control Software4.0Not stated
IlluminaNextSeq 550Dx Control Software4.0 (RUO Mode)Not stated
IlluminaNextSeq 550Dx Operating Software1.0.0 through 1.3.1; 1.3.3Not stated
IlluminaNextSeq 1000/2000 Control Software1.4.1 and earlierNot stated
IlluminaNovaSeq 6000 Control Software1.7 and earlierNot stated
IlluminaNovaSeq Control Software1.8Not stated
IlluminaIscan Firmware4.0.0; 4.0.5Not stated
IlluminaIseq 100 FirmwareNot specified in the recordNot stated
IlluminaMiniseq Firmware2.0 and laterNot stated
IlluminaMiseq Firmware4.0 and laterNot stated
IlluminaMiseqdx Firmware4.0.1 and later; 4.0Not stated
IlluminaNextseq 500 Firmware4.0Not stated
IlluminaNextseq 550 Firmware4.0Not stated
IlluminaNextseq 550dx Firmware1.0.0 through 1.3.1; 1.3.3 and later; 4.0Not stated
IlluminaNextseq 1000 Firmware1.4.1Not stated
IlluminaNextseq 2000 Firmware1.4.1Not stated
IlluminaNovaseq 6000 Firmware1.7 and earlier; 1.8Not stated

Version data comes from the organisation that assigned the CVE. Confirm against the vendor advisory before deciding you are unaffected.

How the severity score breaks down

Attack vector
NetworkCan be exploited remotely over a network, including the internet.
Attack complexity
LowNo special conditions are needed; the attack works reliably.
Privileges required
NoneNo account or login is needed.
User interaction
NoneNo action by a victim is required.
Scope
UnchangedImpact stays within the vulnerable component.
Confidentiality
HighAll data handled by the component can be read.
Integrity
NoneNothing can be modified.
Availability
NoneAvailability is not affected.

Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N. Scored by NVD.

Scoring systemScoreSeverityScored by
CVSS 3.110.0CriticalCNA (hq.dhs.gov)
CVSS 3.17.5HighNVD

Official description

Instruments with Illumina Universal Copy Service v2.x are vulnerable due to binding to an unrestricted IP address. An unauthenticated malicious actor could use UCS to listen on all IP addresses, including those capable of accepting remote communications.

Source: the CVE record as published by its numbering authority, via NVD and CVE.org.

The underlying weakness: Binding to an Unrestricted IP Address

The product assigns the address 0.0.0.0 for a database server, a cloud service/instance, or any computing resource that communicates remotely.

How developers prevent this class of flaw

  • Assign IP addresses that are not 0.0.0.0.
  • Unwanted connections to the configured server may be denied through a firewall or other packet filtering measures.

More Binding to an Unrestricted IP Address vulnerabilities (CWE-1327)

Timeline

  1. CVE record published.
  2. CISA published its SSVC assessment (exploitation: none, automatable: yes, technical impact: total).
  3. Record last changed materially (score, affected products, status or references).

References

Third-party and government advisories

Other vulnerabilities in iScan Control Software

All iScan Control Software vulnerabilities

Spotted an error? Tell us.