CMS8000 Patient Monitor: Hidden Functionality
CVE-2025-0626 is a high-severity hidden functionality vulnerability in CMS8000 Patient Monitor from Contec Health, with a CVSS 3.1 base score of 7.5 out of 10. It can be exploited remotely over the network by an unauthenticated attacker, but only if a user takes part, for example by opening a link or file. Affected CMS8000 Patient Monitor versions: All versions. The record does not name a fixed version, so check the vendor advisory. There is no confirmed exploitation so far. EPSS estimates a 1.14% chance of exploitation within 30 days, higher than 67% of scored CVEs.
High severity with no evidence of exploitation so far.
What to do
- Check which version of CMS8000 Patient Monitor you run against the affected versions below, and install the vendor's latest release.
- Until you can update, restrict network access to the affected service so it is not reachable from untrusted networks.
- Because the attack needs a user to take part, warn users about unexpected links and files related to this product until it is patched.
Affected products and versions
| Vendor | Product | Affected versions | Fixed in |
|---|---|---|---|
| Contec Health | CMS8000 Patient Monitor | All versions | Not stated |
Version data comes from the organisation that assigned the CVE. Confirm against the vendor advisory before deciding you are unaffected.
How the severity score breaks down
- Attack vector
- NetworkCan be exploited remotely over a network, including the internet.
- Attack complexity
- HighSuccess depends on conditions outside the attacker's control.
- Privileges required
- NoneNo account or login is needed.
- User interaction
- RequiredA user must take an action such as opening a file or link.
- Scope
- UnchangedImpact stays within the vulnerable component.
- Confidentiality
- HighAll data handled by the component can be read.
- Integrity
- HighData or code can be modified at will.
- Availability
- HighThe component can be taken down completely.
Vector: CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H. Scored by CNA (hq.dhs.gov).
| Scoring system | Score | Severity | Scored by |
|---|---|---|---|
| CVSS 4.0 | 7.7 | High | CNA (hq.dhs.gov) |
| CVSS 3.1 | 7.5 | High | CNA (hq.dhs.gov) |
Official description
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
Source: the CVE record as published by its numbering authority, via NVD and CVE.org.
The underlying weakness: Hidden Functionality
The product contains functionality that is not documented, not part of the specification, and not accessible through an interface or command sequence that is obvious to the product's users or administrators.
How developers prevent this class of flaw
- Always verify the integrity of the product that is being installed.
Timeline
- CVE record published.
- CISA published its SSVC assessment (exploitation: none, automatable: no, technical impact: total).
- Record last changed materially (score, affected products, status or references).
References
Other references
- cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01
- fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-a…
- bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/
- cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor