CareLink Network: Improper Restriction of Excessive Authentication Attempts
CVE-2025-12995 is a critical-severity improper restriction of excessive authentication attempts vulnerability in CareLink Network from Medtronic, with a CVSS 3.1 base score of 9.8 out of 10. It can be exploited remotely over the network by an unauthenticated attacker, without any action from a user. The record lists version December 4, 2025 as the first without the flaw, so updating CareLink Network to that release or later removes the exposure. There is no confirmed exploitation so far. EPSS estimates a 0.36% chance of exploitation within 30 days, which is low compared with other CVEs.
Critical severity: a successful attack would have severe impact.
What to do
- Update CareLink Network to December 4, 2025 or a later version.
- Read the advisory on medtronic.com for workarounds and exact fixed builds.
- Until you can update, restrict network access to the affected service so it is not reachable from untrusted networks.
Affected products and versions
| Vendor | Product | Affected versions | Fixed in |
|---|---|---|---|
| Medtronic | CareLink Network | before December 4, 2025 | December 4, 2025 |
Version data comes from the organisation that assigned the CVE. Confirm against the vendor advisory before deciding you are unaffected.
How the severity score breaks down
- Attack vector
- NetworkCan be exploited remotely over a network, including the internet.
- Attack complexity
- LowNo special conditions are needed; the attack works reliably.
- Privileges required
- NoneNo account or login is needed.
- User interaction
- NoneNo action by a victim is required.
- Scope
- UnchangedImpact stays within the vulnerable component.
- Confidentiality
- HighAll data handled by the component can be read.
- Integrity
- HighData or code can be modified at will.
- Availability
- HighThe component can be taken down completely.
Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H. Scored by NVD.
| Scoring system | Score | Severity | Scored by |
|---|---|---|---|
| CVSS 3.1 | 8.1 | High | CNA (medtronic.com) |
| CVSS 3.1 | 9.8 | Critical | NVD |
Official description
Medtronic CareLink Network allows an unauthenticated remote attacker to perform a brute force attack on an API endpoint that could be used to determine a valid password under certain circumstances. This issue affects CareLink Network: before December 4, 2025.
Source: the CVE record as published by its numbering authority, via NVD and CVE.org.
The underlying weakness: Improper Restriction of Excessive Authentication Attempts
The product does not implement sufficient measures to prevent multiple failed authentication attempts within a short time frame.
How developers prevent this class of flaw
- Common protection mechanisms include: Disconnecting the user after a small number of failed attempts Implementing a timeout Locking out a targeted account Requiring a computational task on the user's part.
- Use a vetted library or framework that does not allow this weakness to occur or provides constructs that make this weakness easier to avoid. Consider using libraries with authentication capabilities such as OpenSSL or the ESAPI Authenticator.
More Improper Restriction of Excessive Authentication Attempts vulnerabilities (CWE-307)
Timeline
- CVE record published.
- CISA published its SSVC assessment (exploitation: none, automatable: no, technical impact: total).
- Record last changed materially (score, affected products, status or references).
References
Advisories
Recent Improper Restriction of Excessive Authentication Attempts vulnerabilities
- 9.8 Critical severity
- 9.8 Critical severity
- 7.5 High severity
- 7.3 High severity
- 8.1 High severity
- 9.8 Critical severity
Spotted an error? Tell us.