CyberSecBulletin
CVE-2025-14097NVD status: Deferred

ABL90 FLEX and ABL90 FLEX PLUS Analyzers: Improper Authentication

CVE-2025-14097 is a high-severity improper authentication vulnerability in ABL90 FLEX and ABL90 FLEX PLUS Analyzers from Radiometer Medical Aps, with a CVSS 3.1 base score of 7.2 out of 10. It can be exploited remotely over the network by an attacker with administrator privileges, without any action from a user. There is no confirmed exploitation so far. EPSS estimates a 0.46% chance of exploitation within 30 days, which is low compared with other CVEs.

Next patch cycle

High severity with no evidence of exploitation so far.

What to do

  1. Check which version of ABL90 FLEX and ABL90 FLEX PLUS Analyzers you run against the affected versions below, and install the vendor's latest release.

Affected products and versions

VendorProductAffected versionsFixed in
Radiometer Medical ApsABL90 FLEX and ABL90 FLEX PLUS AnalyzersNot specified in the recordNot stated
Radiometer Medical ApsABL800 BASIC and ABL800 FLEX AnalyzersNot specified in the recordNot stated
Radiometer Medical ApsAQT90 FLEX AnalyzersNot specified in the recordNot stated

Version data comes from the organisation that assigned the CVE. Confirm against the vendor advisory before deciding you are unaffected.

How the severity score breaks down

Attack vector
NetworkCan be exploited remotely over a network, including the internet.
Attack complexity
LowNo special conditions are needed; the attack works reliably.
Privileges required
HighAdministrator-level access is required first.
User interaction
NoneNo action by a victim is required.
Scope
UnchangedImpact stays within the vulnerable component.
Confidentiality
HighAll data handled by the component can be read.
Integrity
HighData or code can be modified at will.
Availability
HighThe component can be taken down completely.

Vector: CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H. Scored by CNA.

Official description

A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software. Other related CVE's are CVE-2025-14095 & CVE-2025-14096. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer. Temporary work Around: If the network is not considered secure, please remove the analyzer from the network. Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.

Source: the CVE record as published by its numbering authority, via NVD and CVE.org.

The underlying weakness: Improper Authentication

When an actor claims to have a given identity, the product does not prove or insufficiently proves that the claim is correct.

How developers prevent this class of flaw

  • Use an authentication framework or library such as the OWASP ESAPI Authentication feature.

More Improper Authentication vulnerabilities (CWE-287)

Timeline

  1. CVE record published.
  2. CISA published its SSVC assessment (exploitation: none, automatable: no, technical impact: total).
  3. Record last changed materially (score, affected products, status or references).

References

Other vulnerabilities in ABL90 FLEX and ABL90 FLEX PLUS Analyzers

All ABL90 FLEX and ABL90 FLEX PLUS Analyzers vulnerabilities

Recent Improper Authentication vulnerabilities

Spotted an error? Tell us.