Welch Allyn Connex Spot Monitor: Use of Default Cryptographic Key
CVE-2024-1275 is a critical-severity use of default cryptographic key vulnerability in Welch Allyn Connex Spot Monitor from Baxter, with a CVSS 4.0 base score of 9.1 out of 10. It can be exploited remotely over the network by an unauthenticated attacker, without any action from a user. Affected Welch Allyn Connex Spot Monitor versions: 1.52 and earlier. The record does not name a fixed version, so check the vendor advisory. There is no confirmed exploitation so far. EPSS estimates a 0.39% chance of exploitation within 30 days, which is low compared with other CVEs.
Critical severity: a successful attack would have severe impact.
What to do
- Check which version of Welch Allyn Connex Spot Monitor you run against the affected versions below, and install the vendor's latest release.
- Until you can update, restrict network access to the affected service so it is not reachable from untrusted networks.
Affected products and versions
| Vendor | Product | Affected versions | Fixed in |
|---|---|---|---|
| Baxter | Welch Allyn Connex Spot Monitor | 1.52 and earlier | Not stated |
Version data comes from the organisation that assigned the CVE. Confirm against the vendor advisory before deciding you are unaffected.
How the severity score breaks down
- Attack vector
- NetworkCan be exploited remotely over a network, including the internet.
- Attack complexity
- HighSuccess depends on conditions outside the attacker's control.
- Attack requirements
- NoneThe attack does not depend on a particular deployment or configuration.
- Privileges required
- NoneNo account or login is needed.
- User interaction
- NoneNo action by a victim is required.
- Confidentiality (vulnerable system)
- HighSerious loss of confidentiality on the vulnerable system.
- Integrity (vulnerable system)
- HighSerious loss of integrity on the vulnerable system.
- Availability (vulnerable system)
- NoneNo loss of availability on the vulnerable system.
- Confidentiality (other systems)
- NoneNo loss of confidentiality on systems beyond the vulnerable one.
- Integrity (other systems)
- NoneNo loss of integrity on systems beyond the vulnerable one.
- Availability (other systems)
- NoneNo loss of availability on systems beyond the vulnerable one.
Vector: CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X. Scored by CNA (baxter.com).
Official description
Use of Default Cryptographic Key vulnerability in Baxter Welch Allyn Connex Spot Monitor may allow Configuration/Environment Manipulation.This issue affects Welch Allyn Connex Spot Monitor in all versions prior to 1.52.
Source: the CVE record as published by its numbering authority, via NVD and CVE.org.
The underlying weakness: Use of Default Cryptographic Key
The product uses a default cryptographic key for potentially critical functionality.
How developers prevent this class of flaw
- Prohibit use of default, hard-coded, or other values that do not vary for each installation of the product - especially for separate organizations.
- Force the administrator to change the credential upon installation.
- The product administrator could change the defaults upon installation or during operation.
More Use of Default Cryptographic Key vulnerabilities (CWE-1394)
Timeline
- CVE record published.
- CISA published its SSVC assessment (exploitation: none, automatable: no, technical impact: total).
- Record last changed materially (score, affected products, status or references).
References
Other references
Recent Use of Default Cryptographic Key vulnerabilities
- 9.1 Critical severity
Spotted an error? Tell us.